Care of
the Reference Materials
The blocks
are supplied in a velvet lined case to
protect the fine surface finish produced at manufacture. Providing
that they are handled with care (cotton gloves are recommended)
and always stored in the case
provided, they should give many years of satisfactory service. Exposure
to organic solvents can cause severe fogging of the surface and
must be avoided, as should any attempt to clean the block with abrasive
cleaners.
In the event
of the surface of a block becoming soiled, it should be immersed
in dilute neutral detergent solution which may be warmed to 40°.
Low power ultrasonic vibration may also be used to remove the contamination.
It should then be rinsed thoroughly with deionized water and dried
carefully with a soft lint-free cloth.
|
|
Recorded
Spectra
The following
spectra were recorded in the ratio
mode, using a Perkin Elmer LS50B Luminescence Spectrometer. The
spectral band width for both
excitation and emission monochromators, was 2.5 nm. Data points
were recorded at 0.5 nm intervals. These spectra are specific to
this instrument and the operating conditions used. Other instruments,
or different
operating conditions will produce spectra with maxima which may
vary both in terms of relative magnitude and wavelength.
The corrected
Spectra were recorded using a Perkin Elmer MPF44-B and should relate
to others recorded in a similar manner.
|