Certified Reference Material (RM) Sets

The Certified Reference Material sets have been assembled to make your verification task easier, and meet specific regulatory requirements. In addition, set prices do offer price savings over the purchase of individual Certified Reference Materials.
So where possible use the general syntax rules to construct your own set, even if it is simply an absorbance and wavelength set.
For example, a 60 mg/l potassium dichromate, together with a holmium oxide solution can be ordered under the part number RM-06HL.
Primary Requirement Catalogue Number Construction Parameter Material Wavelength
Full specification verification RM-06HLKCSITX
Absorbance Wavelength Stray light
Resolution
Potassium dichromate 60 mg/l, blank Holmium oxide Potassium chloride, sodium iodide, blank Toluene in Hexane 235 to 350 nm
240 to 650 nm
200 & 260 nm
265 to 270 nm
Photometric linearity
RM-0204060810 Absorbance Potassium dichromate 20, 40, 60, 80, 100 mg/l, 0.001M perchloric acid blank 235 to 350 nm
Pharmacopoeia compliance RM-06HLKCTX Absorbance
Wavelength Stray light Resolution
Potassium dichromate 60 mg/l, blank
Holmium oxide Potassium chloride, blank
Toluene in Hexane
235 to 350 nm
240 to 650 nm
200 nm
265 to 270 n
Similarity to NIST (USA) SRMs
RM-06HLKI Absorbance
Wavelength Stray light
Potassium dichromate 60 mg/l, blank
Holmium oxide Potassium iodide, blank
235 to 350 nm
240 to 650 nm
260 nm
Similarity to NIST (USA) SRMs RM-1N2N3N Visible Absorbance N.D. glass 1.10, 0.78, 0.56 A., blank 440 to 635 nm
Similarity to NIST (USA) SRMs RM-N1N35N Visible Absorbance
N.D. glass 2.10, 1.60, 0.30 A., blank nm 440 to 635
High end visible wavelength RM-06DLKI Absorbance
Wavelength Stray light
Potassium dichromate 60 mg/l, blank
Didymium Potassium iodide, blank
235 to 350 nm
290 to 870 nm
260 nm
  CRM enquiry
Compliance references
Material Test Range USP Ph. Eur. BP ASTM TGA
Potassium Dichromate Absorbance UV/Visible * * * * *
Neutral Density Filters Absorbance Visible * * * * *
Holmium Oxide Wavelength UV/Visible * * * * *
Potassium Chloride Stray Light UV * * * * *
Potassium Iodide Stray Light UV * . . * .
Sodium Iodide Stray Light UV . * * * *
Toluene in Hexane Resolution UV . * * . *
Recertification
How often do I need to recertify my Certified Reference Material?
To ensure the validity of the calibration values of these Certified Reference Material’s it is recommended that they are re-certified at regular intervals. The periodicity of this recerti­fication is under the control of the user, and is specific to the laboratory environment under which the unit is used.

To establish a statistical basis for determining this recertification interval, it is suggested that the Certified Reference Material in question be re-certified every 12 months for at least the first three years of use, and thereafter at a periodicity deemed suitable for the laboratory environment;
the interval between calibrations being established from the data generated in the first three years. Review of this process should be incorporated in any formal control of the system, and any major changes in the laboratory environment should be evaluated as to their impact on the calibration of the unit.     (see MSDS)

Base nav bar REference materials Stray light references Fluorescence References Ordering Recertification Faq's